Results from AgNP method optimization · 2017. 9. 26. · Hg [ng] Ave. Hg/Ag Non-washed AgNPs 67...

Post on 31-Aug-2021

0 views 0 download

transcript

S. Böttgera,b, C. Pipkeb, D. Rosenbergb, W. Jansenb, U. E. A. Fittschena,c

aChemistry Departement, Washington State University, Pullman, WA 99163bDepartement of Chemistry and its teaching methods, Europa-Universität Flensburg, Flensburg, Germany, 24943

cTechnische Universität Clausthal, Institut für Anorganische und Analytische Chemie, Clausthal-Zellerfeld, Germany, 38678

Introduction

Determination of gas phase mercury –Evaluation of internal standards and external calibration

Sebastian Böttger

Sebastian.boettger@uni-flensburg.de

Prof. Dr. Ursula FittschenUrsula.fittschen@tu-clausthal.de

Objective and Experimental

Results from AgNP method optimization

Results from External Calibration

The aim was to develop a reliable, precise, and accurate analysis of gas phaseHg by using TXRF.➢ Chemicals and instrumentation needed are generally already available in

laboratories using TXRF for elemental analysis, and a special setup is thereforenot necessary.

➢ Enrichment of the gas phase Hg on AgNPs was performed and Hg concentrationswere determined using a S2 Picofox TXRF (Bruker, Germany).

➢ A custom built micro-X-ray fluorescence spectrometer (µ-XRF) was used [10] forthe optimization of sample preparation procedures, as well as provide reliableand accurate Hg determination.

➢ AgNPs are produced using AgNO3 and NaBH4. Different procedures ofpreparation and determination were tested: rinsing of deposits after the dryingprocess; absolute determination; internal standards and external calibrationcurves.

Mercury and arsenic compounds were often used to prevent damage tocultural heritage specimens, e.g. herbaria and paintings [1]. Over time, Hg0 airborne is formed. The pathway comprises bacterial activity and redox processesand is still subject of discussion.

Mercury has previously been determined by REXES directly on specimens [1].Portable mercury vapor indicator (MVI) [3,4], SEM, XRD, XPS and ICP-MS [3],and a portable analyzer using AAS [6] have also been used on a variety ofsamples.

By using Total Reflection X-Ray Fluorescence (TXRF), workgroups havedetermined Hg in other samples like wastewater [6], seawater [7], and otherliquid samples using Au-nanoparticles [8], and in seafood samples by using Ag-nanoparticles (AgNPs) [9]. To the best of our knowledge no data is publishedfor the determination of gas phase Hg by using the AgNP-TXRF approach.

Optimization of the Ag-NP preparation method µ-XRF evaluation of internal standards and conclusions about homogeneity

Figure 3: µ-XRF images of elemental distribution. Color intensity ofRGB images is scaled between 5 and 80 % of the maximumnumber of counts. 1st row Ag and Cr (not washed), 2nd row Ag andMo (not washed), 3rd row Ag washed and Mo.

A B

C

D

Figure 2: Different angles for the X-ray beam(Table 1).

Ag/IS and amount of Ag [mg/L]

Total deposit

Ag/IS and Ag [mg/L] Area A

Ag/IS and Ag [mg/L] Area B

Ag/ISand Ag [mg/L] Area C

Ag/IS and Ag [mg/L] Area D

Max. Deviation

from total [%]

Ag and Cr (not washed)

7.92; 113.712.2; 73.8

6.27; 143.6

7.88; 114.3

7.88; 114.2

35

Ag and Mo (not washed)

0.033; 9620.033;

9760.035;

9240.035;

9260.032;

9874

Ag washedand Mo 0.041; 776

0.047;683

0.03;926

0.053;607

0.035;908

22

Table 2: Results relating Ag to two internal standards, Cr & Mo, andthe resulting deviation for different areas.

The Cr-IS is not homogeneouslydistributed on the AgNPs and thereforegives highly variable results dependingon the angle of irradiation (Table 2, Fig. 2,Fig. 3). The Mo-IS is more homogeneousand gives better results than the Cr-IS(Table 2, Fig. 3).

The AgNPs were dropped on a Quartz-carrier and placed in a Hg saturatedatmosphere and allowed to equilibrate for24 h (Fig. 1).

Efficiency and reproducibility of the Hg-capture of washed and non-washed AgNP-specimens were studied (Table 1). Theaverage standard deviation of Ag was ca.10 % within a batch. Washed carrierscontained ~60% less Ag than non-washedcarriers. In addition to lower Ag levels,method optimization also resulted inincreased Hg capture on washed carriers.

References:

[1] SIROIS, Collection Forum, 16 (1-2) (2001), 65-75.

[2] VAUPEL, Deutsches Museum Wiss. Jahrbuch (1989), 189-226.

[3] FELLOWES et al., J. Haz. Mat., 189 (2011), 660-669.

[4] BRIGGS et al., New Phytol., 94 (1983), 453-457.

[5] OYARZUN et al., Science Tot. Environ., 387 (2007), 346-352.

[6] MARGUÍ et al., Talanta, 82 (2010), 821-827.

[7] KOULOURIDAKIS et al., Instr. Science & Techn., 34:4 (2006), 425-433.

[8] BENNUN, GOMEZ, Spetrochimica Acta Part B, 52 (1997), 1195-1200.

[9] ROMERO et al., J. Anal. Spectrom., 29 (2014), 696-706.

[10] FITTSCHEN et al. XRS (2017), sccepted manusscript.

Neither IS gave completely satisfactoryresults. A five-point calibration curve wasused for external calibration. Since AgNPweights on the carriers fell in the range of100 - 200 ng, two calibration ranges wereselected for Ag: 100 – 500 ng and 60 –250 ng.➢ A linear regression analysis was

performed following the calculation ofthe residuum and elimination ofoutliers (Fig. 4).

➢ Higher Ag masses (> 300 ng) resulted inhigher variability between replicates.

Fig. 4: External calibration of Ag.

➢ A five-point calibration range of 60 –200 ng was used for Hg.

➢ A linear regression analysis wasperformed following the calculation ofthe residuum and elimination ofoutliers (Fig. 5).

➢ The results deviate from the absolutevalues measured by the programSpectra.

➢ The maximal allowable work-placeconcentration for Hg is 0.020 mg/m3,which corresponds to ca. 850 pgabsolute on the carrier. Since theactual concentrations in the museums

➢ Lower limits of detection (LLD) for Ag and Hg were found to be 129 pg and 4pg, respectively.

➢ The calibration of Hg needs to be performed in a significantly lowerconcentration range. Since the maximum allowable work-place Hgconcentration is 0.020 mg/m3, and concentrations found in the Flensburgmuseum, as well as other museums, a calibration range from 50 pg to 1 ngwould be favorable. This would also significantly reduce matrix effects.

y = 310.1x + 6598.4R² = 0.8121

0

50000

100000

150000

200000

250000

0 100 200 300 400 500 600

Net

Co

un

ts A

g (L

1)

Mass Ag (ng)

External Calibration (Ag) Linear (External Calibration (Ag))

y = 1777.9x + 135789R² = 0.5649

0

100000

200000

300000

400000

500000

600000

700000

0 50 100 150 200 250

Net

Co

un

ts H

g (L

1)

Mass Hg (ng)External Calibration (Hg) Linear (External Calibration (Hg))

Fig. 5: External calibration of Hg.

TreatmentAve.

Absolute Ag [ng]

Ave. Absolute Hg [ng]

Ave. Hg/Ag

Non-washedAgNPs

67 0.5354 0.00085

WashedAgNPs

45 6.09396 0.01379

Table 1: Results relating absolute

weights of Hg and Ag for washed

and non-washed AgNPs.

Fig. 1: Quartz-carrier with 10 µL of the AgNP allowed to equilibrate for 24 h in an Hg-saturated atmosphere.

Ag L IS Kα RGB image

Conclusions were found to be around 0.002 mg/m3 (ca. 100 pg on the carrier), acalibration curve range of 0.05 – 1 ng would be more suitable forquantification purposes. A lower calibration curve range is also expected tobe less prone to errors due to absorption effects.

Acknowledgements: The authors would like to thank the Fittschen Group andZentraler Ausschuss für Europa und Internationales der Europa-UniversitätFlensburg for a travel grant to do research at Washington State University, as wellas the Ausschuss für Forschung und Entwicklung and the Abteilung Chemie undIhre Didaktik der EUF.